Operation
Setting/releasing footbrakes
Warning
Surgical microscope can move without warning!
➩Always set footbrakes, except during transport.
Setting footbrakes
➩Loosen star grip (2).
➩Turn lever for footbrake (1) to the right.
➩Tighten star grip again.
1
Releasing footbrakes
➩Loosen star grip (2).
➩Turn lever for footbrake (1) to the left.
➩Tighten star grip again.
14
2
2
1
Leica M500 OH3 / Ref. 10 711 894 / Version -
Locking the Leica M500 OH3
The D axis of the Leica M500 OH3 is locked when locking is
performed.
Locking is mainly used for installation and accessory-change at
the Leica M500 OH3.
➩Pull out and turn locking knobs (4) until both points (arrows)
are aligned.
➩Move swing arm up and down until the lock (5) engages.
➩Move swing arm forward and backward until the lock (3)
engages.