Operation
➩When the audible signal stops, press the ALL-FREE button
and swivel the microscope carrier into position B (arrow)
until it engages.
If the mounted accessories (e.g. the tube) do not permit
the 90° swivel movement, turn the tube upward,
perform the swivel movement and then ensure that you
turn the tube back to the working position.
➩Lift the protective flap and press the pushbutton (Item 3, Page
15) again for auto balance.
The pushbutton flashes orange and an audible signal is
emitted for the entire duration of the balancing procedure for
the B, C and D axes.
Balancing is complete when the pushbutton for auto balance
illuminates green and none of the three LEDs (Item 4, 5 or 6,
Page 15) illuminates.
If one of the three LEDs illuminates, balancing must be
checked:
➩Press the ALL-FREE button and check balancing.
Correct balancing manually if necessary.
16
Correcting balancing manually
➩Press the ALL-FREE button.
The microscope tilts to the side:
➩Press the C (1) button until the microscope is balanced out.
The microscope swivels to the front or back:
➩Press the A (2) button until the microscope is balanced out.
Leica M500 OH3 / Ref. 10 711 894 / Version -
Ensure that no accessory parts collide with the
microscope when balancing manually.
As the LEDs (Item 4, 5 and 6, Page 15) also show very
slight balance deviations, the manual balance
correction may only be performed for perceptible
deviations.