Leica M500 OH3 User Manual page 10

Hide thumbs Also See for M500 OH3:
Table of Contents

Advertisement

Introduction
Illumination
The illumination of the Leica M500 OH3 surgical microscope
consists of a main illuminator and a backup illuminator.
They are located in the illumination unit.
The main and backup illuminators consist of xenon lamps. The
light of the main and backup illuminators is lead to the optics
carrier via a fibre-optic cable through the base.
If the main illuminator fails, the plug of the fibre-optic cable is
switched over to the backup illumination.
10
1 Access hatch for lamp inserts of main and backup
illuminators
2 Regulating knob for brightness of main illuminator
3 Outlet for fibre-optic light guide of main illuminator
4 Regulating knob for brightness of backup illuminator
5 Outlet for fibre-optic light guide of backup illuminator
6 Power switch of Leica M500 OH3 surgical microscope
7 100 V/120 V/220 V/240 V switch
8 Power connection
9 Pullout fuse for power cable
10 Potential equalisation connector
8
1
Leica M500 OH3 / Ref. 10 711 894 / Version -
Two LEDs at the control unit indicate which illuminator is being
used at the moment (see Page 18).
The light field diameter can be concentrated at the optics
carrier with the rotary knob (11).
2
11
3
4
5
6
7
8
9