Precautions for Exposure
1.2 Depiction of the Cervical Region
1 The radiation field must not include the whole head. Be sure to secure transparent
portions on both sides of the neck.
Use the "Skull, General" menu to include the whole head in the radiation field.
2 For exposure of the pharynx and/or larynx, be sure that the neck comes to the center
of the radiation field, especially in terms of upper and lower margins.
3 In pharynx and/or larynx exposure, do not use lead characters in the oblique line
;;
section.
;;
1.3 Depiction of HIP JOINT AXL - 2
1 Make sure to position the region of interest to the slanted-line area shown below. Do
not collimate further inside.
2 Positioning should be done so that the condyle and the femur run along the longer
edge. (Do not have them positioned against the shorter edge.)
Condyle side
Femur side
Z-4
( )
( )
;;
;;
1/4
1/4
7/12
1/12