Terminate testing

7 Press the Home navigation key, and then restart the test.
This will clear all alarms and begin a new test.
8 Press the Display soft key, and then select Signal.
The Signal Results window appears.
9 Check the timing slips result.
For a description of the result, see
page
124.
You have finished measuring timing slips.
In Terminate mode, you can use the HST-3000 to perform bit
error rate testing (BERT or BER testing) on E1 circuits, and
timeslots within the E1 circuit. You can test for bit errors, code
errors, frame errors, and CRC errors (if applicable).
In Terminate mode, it is assumed that there is either a far-end
loopback or another test set terminating the far end. This
allows you to qualify E1 circuit error performance by testing for
bit errors, code errors, CRC errors, FAS errors, MFAS errors,
and E-BIT errors (if applicable) on E1 lines.
The following procedure describes how to perform a BER test.
To perform a BER test
1 Launch the BERT application. See
cation" on page
HST-3000 E1 Testing User's Guide
5.
Chapter 2 E1 testing

Terminate testing

"Timing Slips" on
"Launching an appli-
27