During the calibration process outlined in Step 7, the user may wish to verify
the correction point that is to be programmed next. This can be accomplished
key. The LCD will display CP X, where X is
by Momentarily pressing the
F3
the next correction point number to be programmed. This feature is helpful
after programming many correction points since it can be easy to lose track of
the number of standards programmed.
NOTE: If an error is made during the calibration process where an incorrect
table entry is recorded, the entire process must be repeated.

Operating in Compensated mode.

After a successful correction table has been programmed, the Readout will
operate normally and the measuring system will have a higher level of
accuracy.
NOTE: If the encoder is positioned outside the calibrated (corrected) range of
operation, the LCD will display No Co alternately with the current position.
This is normal and indicates that the system is operating outside the calibrated
measuring range and has a reduced measuring accuracy.
If you choose to turn off Non-Linear Compensation (
) the compensation
Pr39=0,
table will be saved in memory. Thereafter, changing
to 1 will re-enable the
Pr39
previously stored compensation table.
NOTE:
Resetting
the Readout to factory default settings does not erase the
compensation table.