External Power

For 12-24VDC Power INPUT connections, see
this manual.

Output Signal Configuration

Set
JP5
on the readout circuit board to position A.
Set
P
ROGRAMMING
signal.

Output Signal Polarity

The polarity of the output signal is set using
The output is a transistor that conducts to Negative. The factory default is
normally open (NO).
NOTES: The LCD will flash when any of the programmed conditions above are
encountered, but the output signal only changes state once and then toggles
back when an in-tolerance condition is restored.
When the Monitor mode is active AND the programmable output is enabled, the
hardware output will not become active until the error condition has lasted at
least 2 consecutive seconds.
CAUTION: The output signal remains active during Programming.
If parameters relating to the output are changed during Programming, the
output signal could become active!
Manual Part # 800-1402-002 Rev K, 072418