Advanced Programming
Functions
Advanced Programming Functions allow the Readout to perform special tasks
or invoke custom programming. These tasks interact with the
functions related to a specific task or programming operation.
Note: The Advanced Functions configurable via
each other. The table below provides a summary of the functions available.
Programming Parameter Pr24 Value
Note: When
P
ROGRAMMING
disabled and have no function. However, the
Addition, is independently controlled by Pr30.
Set Programming Parameter

Monitor

The Readout can monitor a position to detect drift or
measurement variance. To activate the Monitor mode, position
the Encoder (system) to the desired location and Momentarily
press the
key. The MON symbol will illuminate on the LCD to
F3
indicate Monitor mode is active.
If the Encoder moves outside the programmed tolerance the
LCD reading flashes, indicating a drift condition. When the Encoder is moved
back within the programmed tolerance, the displayed reading will stop flashing.
To exit the monitor mode, Momentarily press the
turn off and the currently displayed position will stop flashing.
Monitor can only be activated while in the
key is depressed, or the normal operational mode is changed (enter
programming mode or sending a data transmission), the Readout will
automatically exit Monitor mode.
NOTE: When the Monitor mode is enabled AND the
configured for drift monitoring, the output will not become active until the
system has remained 'out of position' for at least 2 seconds.
Operation Manual for LCD Digital Readouts
0
1
2
3
4
P
Pr24
ARAMETER
= 1 to enable this function.
Pr24
are mutually exclusive of
Pr24
Function
No Advanced Functions.
F3 = MON F4 = HOLD F2 = SPC delete
Go/No Go gauge operation
Measurement Accumulation
Statistics Mode
is set to 0, the
F2, F3 & F4
key, used only for
F1
key. The MON symbol will
F3
ABS measuring
programmable output
keys to perform
F
keys are
Offset
mode. If the
ABS
Page 24 of 53
/