8 Adjustment FC

After a certain period of use, the force gauge may show deviations in a measuring
range which are due to the functioning of the device or other external influences.
In such a case, the device can be sent to our customer service for expert testing and
recalibration.
However, if you have standard force gauges and a measuring stand, you can perform
the calibration yourself according to the instructions below:
1. Fasten the force gauge to the measuring stand or to another holder
2.
Zero the tare value by pressing the key.
3. Call up the calibration menu.
4. Load with test weight. The current measured value now equals the test weight
load. Wait until the measured value stabilizes before reading the measured
value.
5. Use
the and
6. Press
the key to initiate a new calibration procedure. The calibration process
can be interrupted
If the calibration procedure is completed or interrupted three times, a message
window appears with the request to confirm the displayed message "Save and
Exit" (YES) or (NO).
Press
the or
If the "YES" option is selected, the display shows "Calibration complete!
14
keys to enter the test weight.
by pressing the key.
key to select the desired option, then press the key
❶Kalibrierungsdauer
❷Aktueller Measured value
❸Standardwert, entered
FC-BA-e-2020
.