Creating and editing a program:
Tend
hh:mm
1
35.00
2
40.00
00:10
3
50.00
00:20
4
70.00
00:50
 NEW
ESC
Pmp
S1
1
1
on
2
1
on
3
2
on
4
2
on
 OK
ESC
You can quit the edit window at any time without changes using
number, using
Note: No time specification is possible in the start segment. The temperature of the first segment is attained as quickly as
possible in order to switch to segment 2 after reaching the set tolerance.
The programmer edit window contains the following parameters:
The segment number of the program is located in the first column.
Tend:
Final temperature to be attained
hh:mm:
Time in hours (hh) and minutes (mm) in which the specified temperature is to be attained.
If the value "0:00" is entered in the field "hh:mm", the set value is accepted immediately
and the bath temperature approached as quickly as possible.
Tol.:
Defines how exactly the final temperature is to be attained before the next segment it processed.
If the tolerance range is selected too small in the "Tol." field, the program might not continue, because the
required tolerance is not achieved.
Pmp:
Pump level at which the segment is to be processed.
S1, S2, S3:
Switching contacts of the contact module (if present) can be programmed here. Contact modules are availa-
ble as accessories ( 9). The setting "- -" stands for no change to the preceding segment, i.e. if "- -" is pre-
sent in all fields, the contact setting of the start setup or that before the program start is retained.
Inserting a new segment
Tend
1
35.00
2
40.00
3
50.00
4
70.00
 new
ESC
Deleting a segment
10/2019 i
Tol.
--:--
0.1
0.1
0.0
0.1
DELETE
S2
S3
- - -
off
- - -
off
- - -
off
- - -
off
- - -
you return to the menu level of the programmer without changes.
hh:mm
Tol.
--:--
0.1
00:10
0.1
00:20
0.0
00:50
0.1
DELETE
Compare the programming example ( 11.8)
You access the editor view of the programmer by selecting and
confirming 
Programmer
dow information go to the right with .
With the keys , ,
segment fields.
If the cursor is located in the first column, the functions "new" and
"delete" are visible in the soft-key bar. Program steps can be creat-
ed or deleted in this way.
With all other cursor positions the function "OK" is visible in the
soft-key bar.
Using
(OK) you can select the appropriate parameters and
change them with
(ESC). When the cursor is located on a segment
With
or
go to the segment number under which the new seg-
ment is to be inserted.
The new segment is inserted on pressing
described above.
ECO Silver
 Edit. To view the complete win-
and
you obtain access to the individual
or .
(new). You can edit it as
95 / 124