Assembly as cooling thermostat
Notice
10/2019 i
Falling / toppling equipment
Property damage
Do not tilt the cooling device during transport and never turn
it upside down.
After transport, site the device in place where possible two hours be-
–
fore putting it into operation so that, if necessary, oil deposits can
form again and the compressor can develop its maximum power.
Do not cover the ventilation openings.
–
Keep a distance of at least 40 cm free on all sides of the device.
–
Set the flow distribution to INT so that during operation as a bath
–
thermostat (without external consumer) the flow is discharged from
the opening for the internal bath circulation.
Plug the appliance connector of the cooling underpart into the appro-
–
priate socket 51H and the control cable into the connection socket
at the back of the operating panel.
During operation as a bath thermostat without an external consumer
–
and with the pump connection set fitted, the outflow nozzle of the
pump connection set must be closed (use sealing plug) or connected
to the return nozzle with a hose.
For bath temperatures above 70 °C attach the sticker included in the
–
supplied items to an easily visible point on the bath.
Operation with external consumer ( 6.2).
–
ECO Silver
.