IN_DO_01
IN_DO_02
IN_DO_03
IN_MODE_00
IN_MODE_01
IN_MODE_02
IN_MODE_03
IN_MODE_04
TYPE
VERSION_R
VERSION_S
VERSION_B
VERSION_T
VERSION_A
VERSION_V
VERSION_D
VERSION_M_0
VERSION_M_3
VERSION_M_4
VERSION_M_5
VERSION_E
STATUS
STAT
RMP_IN_00_XXX
RMP_IN_01
RMP_IN_02
RMP_IN_03
RMP_IN_04
RMP_IN_05
116 / 124
Command
Status of Contact Output 1:
0 = NO contact open/ 1 = NO contact closed.
Status of Contact Output 2:
0 = NO contact open/ 1 = NO contact closed.
Status of Contact Output 3:
0 = NO contact open/ 1 = NO contact closed.
Keypad: 0 = released / 1 = locked.
Control: 0 = int. / 1 = ext. Pt100 / 2 = ext. Analog / 3 = ext. Serial.
Standby operation: 0 = Device ON / 1 = Device OFF.
Keypad remote control unit Command: 0 = released / 1 = locked.
Setpoint offset source: 0 = normal / 1 = ext. Pt / 2 = ext. Analog / 3 = ext. Serial.
Query of the device type (response = "ECO")
Query of the software version number of the control system.
Query of the software version number of the protection system.
Query of the software version number of the Command remote control.
Query of the software version number of the cooling system.
Query of the software version number of the analog module.
Query of the software version number of the RS 232/485 module.
Query of the software version number of the digital module.
Query of the software version number of the solenoid valve (cooling water).
Query of the software version number of the solenoid valve (shut-off valve 1).
Query of the software version number of the solenoid valve (shut-off valve 2).
Query of the software version number of the high temperature cooler.
Query of the software version number of the external Pt100 module.
Query of the device status 0 = OK, -1 = Error.
Query of the error diagnosis response:
XXXXXXX  X = 0 no error, X = 1 error
1st character = Error
2nd character = Alarm
3rd character = Warning
4th character = Overtemperature
5th character = Low Level
6th character = 0
7th character = External control value missing
Query of a program segment XXX
(Response: e.g. 030.00_00010.00_005.00_001.00 => Set-point temperature =
30.00 °C, Time = 10 min, Tolerance = 5.00 K, Pump stage = 1).
Query of the current segment number.
Query of the set program loops.
Query of the current program loops.
Query of to which program further commands refer.
Query of which program is currently running (0 = none).
ECO Silver
Meaning
10/2019 i