Caution!
Caution!
Notice
44 / 124
Bath overflow due to thermal expansion or
immersion of objects
Burns, scalds, frostbite
Take the volume of external consumers into account.
Take into account the increase in volume with a rise in tem-
perature.
Hot vapor formation / discharge of boiling cooling-water on the cool-
Filling of cooling coil with cooling water only admissible up to
T
of 100 °C!
max
Inadmissible operating temperatures;
temperature difference between outflow and product
Property damage (consumers, external components)
Note that an externally controlled bath temperature, espe-
cially during a transient response, may differ substantially
from the set-point temperature.
Note the various limitation options (Tih, Til, T
limitation).
Set the overtemperature switch-off point T
the heat transfer liquid. T
ECO Silver
ing coil
Burns, scalds
too large
max
must be below the flash point.
max
, correction
max
according to
10/2019 i