4.0 Operation
Fig. 15.
Fig. 23.
Fig. 24.
4.8

Insert DDS hose adapter

Press the required DDS hose adapter with 6 or 10 mm
diameter into the "Patient" hole of the DDS collection jar
lid twisting slightly to ensure a tight fi t.
Twist slightly in the same manner when removing.
4.8

Connect hose

Fix the hose tightly to the adapter (olive).
4.9

On/off switch

Press the "I" symbol to switch the unit on.
Press the "0" symbol to switch the unit off.
The unit is ready for optional footswitch operation in the
„0" position (see options).
4.10

Set vacuum

Close the suction hose and set the desired vacuum by
turning the vacuum controller according to the direction of
the arrow.
Do not use force to turn the knob at its limits!
Test the system for leaks if the desired vacuum is not
reached. (Please also see chapter 8.0 Trouble shooting)