Operation
10.2 Parameter for the measurement
To get correct readings, the parameters
• Existing pressure
• Distance to measuring point
are needed.
Home
Configuration
or
Setting buttone
Note:
Distance refers to distance from measuring point to the measurement tool.
Fixed values are defined for the focus tube and the gooseneck, i.e. 20 cm and 5 cm.
Permissible input range for the acoustic trumpet is 1m to 6m, for the parabolic mirror
3m to 12m.
Pressure input range is 0.3bar to max. 10bar, for vacuum leakages -0.1bar to -1bar.
Testo Sensor LD pro V2.00
Parameter
Pressure/ Distance
Parameter
Pressure/ Distance
The pressure entry can be made by selecting one
of the 3 predefined values or via the text field.
Max. Permissible pressure value is 10bar.
In case of higher pressures, please insert 10bar)
For the distance input, acoustic trumpet /
parabolic mirror, there are 3 predefined values
or a text field for the measuring-specific distance
input.
Acoustic trumpet:
3 - 6m
Parabolic mirror:
3-12m
Page 18 of 48