Installation In Hazardous Location - Siemens SITRANS F M MAG 5000 Manual

Hide thumbs Also See for SITRANS F M MAG 5000:
Table of Contents

Advertisement

Safety notes

2.2 Installation in hazardous location

CE-marked equipment
2.2
Installation in hazardous location
Ex approvals
Temperature specifications for Ex use
Temperature class
T2
T3
T4
T5
T6
EX requirements
12
www.eltra-trade.com
The CE mark symbolizes the compliance of the device with the following guidelines:
● EMC-guideline 89/336/EEC
● Low voltage guideline 73/23/EWG
● ATEX Directive 94/9/EG
● CT: (MI-001) Directive 2004/22/EC
WARNING
Equipment used in hazardous locations must be Ex-approved and marked accordingly.
It is required that the special conditions for safe use provided in the manual and in the Ex
certificate are followed!
CSA Class I, Division 2, Groups A, B, C and D. Code T5 for an ambient temperature of
+60 °C.
FM Class I, Division 2, Groups A, B, C and D and Class I, Zone 2, Group IIC indoor/outdoor
Type IP67 hazardous (classified) locations
Ambient temperature [°C]
-40 to +40
180 (process temperature)
165 (process temperature)
100 (process temperature)
65 (process temperature)
50 (process temperature)
It is required that:
● Electrical connections are in accordance with Elex V (VO in explosion hazardous areas)
and EN60079-14 (Installing Electrical Systems in Explosion Hazardous Areas).
● The protective cover over the power supply is properly installed. For intrinsically safe
circuits the connection area can be opened.
-40 to +50
-
140 (process temperature)
100 (process temperature)
65 (process temperature)
50 (process temperature)
info@eltra-trade.com
-40 to +60
-
-
80 (process temperature)
65 (process temperature)
50 (process temperature)
+421 552 601 099

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents