Siemens SITRANS F M MAG 5000 Manual page 65

Hide thumbs Also See for SITRANS F M MAG 5000:
Table of Contents

Advertisement

Troubleshooting/FAQs
Symptom
Measuring approx.
50%
Loss of totalizer data
##### Signs in
display
62
www.eltra-trade.com
Output
Error
Cause
signals
code
F63
Defective SENSORPROM
unit
F71
Loss of internal data
Maximum
W30
Flow exceeds 100% of Qmax.
W21
Pulse overflow
Volume/pulse too small
Pulse width too large
Missing one electrode
connection
OK
W20
Initializing error
OK
Totalizer roll over
®
info@eltra-trade.com
Remedy
Replace SENSORPROM
unit
®
Replace MAG 5000/6000
Check Q
(Basic Settings)
max
Change volume/pulse
Change pulse width
Check cables
Reset totalizer manually
Reset totalizer or increase totalizer
unit
+421 552 601 099