7.4 Reset mode
Resetting
Zero point adjustment (MAG 6000 SV only)
50
www.eltra-trade.com
1. Press top up key
for 2 sec.
2. Enter password.
3. Use forward key
or backward key
4. Press lock/unlock key
5. Press forward key
to reach totalizer/counter to be reset or default setting menu.
6. Press lock/unlock key
If restoring of factory settings is required:
1. Press lock/unlock key
See also reset menu diagram.
Auto adjustment
Before auto zero point adjustment is carried out ensure that valves to and from flowmeter are
completely closed and that flow velocity in sensor is zero.
1. Press top up key
for 2 sec.
2. Enter password.
3. Use forward key
or backward key
4. Press lock/unlock key
5. Press forward key
to reach zero adjust menu.
6. Press lock/unlock key
7. Use change key
to select "auto".
8. Press forward key
to view actual offset (lower line in display). Value will be zero after
adjustment has been performed.
9. Press lock/unlock key
Manual adjustment.
1. Press top up key
for 2 sec.
2. Enter password.
3. Use forward key
or backward key
4. Press lock/unlock key
5. Press forward key
to reach zero adjust menu.
6. Press lock/unlock key
7. Use change key
to select "manual".
to reach reset mode menu.
to enter reset menu.
to start resetting.
again to confirm destruction of customized settings.
to reach reset mode menu.
to enter reset menu.
to enter the menu.
to start adjustment.
to reach reset mode menu.
to enter reset menu.
to enter the menu.