The REMOTE menu also displays the type of laser selected. Do not change this field.

Display Option

When the DISPLAY button on the main menu is pressed, some important system parameters
for the SOM are displayed, as shown in Figure 50. The parameters include:
Serial number of the SOM
Number of points in a trace. The number can be used to set the MaxPoints in the
SWS15100.INI file.
Value of M
, one of the wavelength calibration constants
O
Value of L
, the other wavelength calibration constant
O
The other constants displayed are currently not used.

Source Optics Module Operation

During SOM operation, the display is updated to show the current operating conditions. When
the laser is commanded to scan, dots are added to the bottom line of the screen, which creates
a moving display, as shown in Figure 51. The display indicates that the SOM is operative. If the
dots do not appear, or do not change, the SOM is not operating correctly. Error messages are
displayed on the same line and remain on the screen until a button is pressed.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Figure 49: Remote Option on the SOM
Figure 50: Display Option on the SOM
Operating and Maintenance Instructions – 63