Figure 91. Click "YES" to restore the last calibration data
Note. Latest calibration data currently stored in the EEPROM is lost and the old calibration data
is restored.
If there is more than one receiver chassis hooked up to your transmitter, use the
instructions above to calibrate one receiver chassis, and then copy the calibration
constants to the Control Modules in the other receivers. This saves you time.

Calibrating Multiple Receivers

Up to four receiver stations can be connected to each transmitter station. To copy the calibration
constants from the first receiver to other receivers:
1. Perform the wavelength calibration on one receiver. Save the new calibration data, but do not
save the old calibration.
2. Save the calibration data again by selecting as if to save the old calibration data. This
procedure saves the current calibration data in the SWS15100.INI file in the C:\WINDOWS
directory.
3. Copy the SWS15100.INI file to a diskette.
4. At each additional receiver station, edit the SWS15100.INI file, using a text editor such as
Notepad, by copying the old calibration data from the diskette to the file. Save the new
SWS15100.INI file as text.
5. Run the SWS program.
6. From the main window of the software, select the System menu, choose the Calibration
option, and select the Wavelength Calibration tab. The Wavelength Calibration window
appears.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Operating and Maintenance Instructions – 109