Enabling Wavelength Diagnostics On C-Band And L-Band Sws Systems - JDS Uniphase SWS15100 User Manual

Swept wavelength systems
Table of Contents

Advertisement

1. From the Power Calibration window (Figure 95), select one channel from the Selected
Channels list, click the Process button, and select the Set to Factory Settings option. The
factory settings are reset.
2. Repeat the process for the next channel.
To restore the factory wavelength calibration:
1. In the SWS program, select Insertion Loss mode.
2. From the System menu, select Calibration.
3. Select the Wavelength Calibration tab in the dialog box that appears.
4. Click the Restore Calibration button at the bottom of the dialog box.
5. In the message box that appears, click the No button to select Factory Calibrated Values.
If there are unexpected results, contact JDS Uniphase.

Enabling Wavelength Diagnostics on C-Band and L-Band SWS Systems

In systems used in applications where it is necessary to monitor wavelength accuracy, or where
wavelength instability is suspected, a wavelength diagnostic feature can be enabled. When the
feature is enabled, the wavelength accuracy is tested after every scan and errors are logged to a
file. The user is informed immediately of any wavelength instability.
To enable wavelength diagnostics on both C- and L-band systems:
1. Edit the SWS15100.INI file.
2. Reset the number of points on the SOM.
3. Perform wavelength calibration.
118 – Operating and Maintenance Instructions