Installing the Compact Two-Channel Receiver

The compact two-channel receiver is an optional substitute for, or add-on to, the larger
OWB10002 chassis-based receiver. The two-channel receiver can be installed only as the first
receiver in a daisy-chained series of receiver chassis. To install the receiver:
1. Connect the supplied parallel port cable to the parallel port on the computer and to the
parallel port connector on the two-channel receiver.
2. If the receiver is being daisy-chained, connect the 50-pin ribbon cable to the connector on the
back panel of the two-channel receiver.
3. Connect the other end of the 50-pin ribbon cable to the uppermost 50-pin connector on the
OWB10002 chassis.
4. Plug the unit into a power socket, ensuring that the ground for this receiver is the same
ground as for the other receiver modules and the computer.
5. The optical connections to the two-channel receiver are on the front panel. Connect the
output of the source optics module to the input of the receiver (left-most FC/APC connector).
6. The output of the control module is located beside the input of the receiver. Use the supplied
FC/PC-FC/APC 1 m jumper to connect the output of the control module to Channel 1 or
Channel 2.
7. Connect the other end of the FC/APC cable to the Control Module output.
8. The receiver cannot be tested at this point. Proceed with the rest of the installation, including
installing the software, as explained in the following sections.

Installing the Data Acquisition Board

The computer system can use the optional data acquisition board instead of the parallel
interface. To install the data acquisition board:
1. Configure the data acquisition board:
a) Ensure the computer is powered down.
b) Remove the data acquisition board from its ESD bag.
c) Verify the settings of the DIP switches and jumpers on the board. Set the DIP switches
settings as shown in Figure 19.