Error Messages in SWS

Error Messages from the Tunable Laser Source

The error messages generated by the tunable laser source are listed in Table 23, with actions to
take when the messages appear.
Table 23: Tunable Laser Source Error Messages
Error Message
WARNING: User calib.
err
Press "Enter"
ERROR: no factory
calib.
Turn power off
DEFAULT
Laser head
disconnected
Value Error
Limit Switch
Turn power off
EEPROM error
Turn power off
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Measurement
Action: Measure power to
the device, with the system
operating.
What is the maximum
power input to the DUT?
Significance
The user calibration is not
correct.
There is no factory
calibration.
There is no connection
between the head and the
controller.
The entered value is not in
the valid range.
There is a mechanical
problem.
There is an electrical
problem.
Answer
Comment
Action
The default value is used instead of
the user calibration.
Switch off the system and check the
electrical connections. Power on. If
the problem persists, call the service
hotline.
Switch off the system and check the
electrical connections. Power on. If
the problem persists, call the service
hotline.
Press Enter. If the system is under
remote control, check the interface
connection and retry. If the error
persists, the value is wrong. Enter a
valid value.
Switch the laser off and call the
service hotline.
Switch the system off and call the
service hotline.
Reference – 215