SWS Performance Checklist

Component
Condition
Laser
During
power-up
System
running
System
stopped
System
stopped
System
stopped
System
stopped
Polarization
Controller
System
running
Source Optics
Module
System
running
System
running
System
running
210 – Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Measurement
What is the laser model
number?
Does the laser stay in
"Initializing"?
Is the laser scanning
properly?
What is the wavelength
error at the start
wavelength?
What is the wavelength
error at the stop
wavelength?
What is the power shown
on the laser display?
What is the laser power
measured at the FC/PC
end of the blue fiber?
Is a polarization controller
connected to the SOM?
What is the polarization
controller model number?
Are the digits on the
display changing with each
scan?
What is the SOM model
number?
What is the maximum
power measured at the
Signal Cond.Out port?
Action: Reconnect Signal
Cond.Out.
What is the maximum
power measured at output
port 1?
Action: Reconnect Signal
Cond.Out.
Answer
Comment
SWS15101
SWS16101
YES
NO
YES
NO
YES
NO
N/A
SWS15104
SWS16104
YES
NO
SWS15102
SWS16102

Hide quick links: