Safety Information, Instructions, and Symbols
Safety Information
Classification
The equipment consists of an exposed metal chassis that is connected directly to earth via a
power cord and, therefore, is classified as a Class 1 instrument. Class 1 refers to equipment
relying on ground protection as a means of shock protection.
The following symbol is used to indicate a protective conductor terminal.
Disconnecting from Line Power
Some of the circuits are powered whenever the units are connected to the AC power source
(line power). To ensure that a unit is not connected to the line power, disconnect the power cord
from either the power inlet on the unit's rear panel or from the AC line-power source (receptacle).
The power cord must always be accessible from one of these points. If a unit is installed in a
cabinet, the operator must be able to disconnect the unit from the line power by the system's
line-power switch.
Line Power Requirements
The equipment can operate from any single-phase AC power source that supplies between 100
and 240 V at a frequency range of 50 to 60 Hz. The typical maximum power consumption is
1500 W.
Fuse Type
The fuse type for the receiver is T2A/250 V. The fuse type for the source optics module is
T1A/250 V. The fuse type for the tunable laser source is T4AL/250 V.
Laser Specifications
Laser specifications are outlined in Table 1.
Parameter
Wavelength
Class
Output power
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 1: Laser Specifications
1520
to
1570
nm
1540
to
1630
nm
1310 nm (source optics module)
I
6 dBm or 4 mW (SWS15101), 3 dBm or 2 mW (SWS16101), and
-3 dBm or 0.5 mW (SWS15102 or SWS16102)
Specification
(SWS15101
laser
source,
(SWS16101
laser
source,
C-band)
L-band)
Safety – 5