Section 5: Operation and Maintenance
March 2014
5.7.1
Switching HART revision with generic menu
If the HART configuration tool is not capable of communicating with a HART Revision 7 device, it
should load a Generic Menu with limited capability. The following procedures allow for
switching between HART Revision 7 and HART Revision 5 from a Generic Menu.
1.
a.
b. To change to HART Revision 7, Enter: HART7 in the message field.
5.7.2
Switching HART revision with field communicator
From the HOME screen, enter the Fast Key sequence and follow steps within the Field
Communicator to complete the HART revision change.
From the HOME screen, enter the Fast Key sequence
Device Dashboard Fast Keys
5.7.3
Switching HART revision with AMS Device Manager
1.
2.
Note
AMS Device Manager versions 10.5 or greater are compatible with HART Revision 7.
5.7.4
Switching HART revision with local operator interface
Navigate to HART REV within the extended menu and select either HART REV 5 or HART REV 7. Use
Figure 5-8
Figure 5-8. Change HART revision with local operator interface
62
Locate "Message" field.
To change to HART Revision 5, Enter: HART5 in the message field.
Click on Manual Setup and select HART.
Select Change HART Revision, then follow the on screen prompts.
below to change HART Revision.
VIEW CONFIG
ZERO TRIM
UNITS
RERANGE
LOOP TEST
DISPLAY
EXTENDED MENU
EXTENDED MENU
EXIT MENU
HART5
2, 2, 5, 2, 4
EXTENDED MENU
CALIBRAT
DAMPING
TRANSFER FUNCT
SCALED VARIAB
ASSIGN PV
TAG
ALARM SAT VALUES
PASSWORD
SIMULATE
HART REV
HART REV
BACK TO MENU
EXIT MENU
Operation and maintenance
Reference Manual
00809-0100-4108, Rev BA
HART7
2, 2, 5, 2, 3