Transient protection terminal block grounding
The transmitter can withstand electrical transients of the energy level usually encountered in
static discharges or induced switching transients. However, high-energy transients, such as
those induced in wiring from nearby lightning strikes, can damage the transmitter.
The transient protection terminal block can be ordered as an installed option (Option Code T1)
or as a spare part to retrofit existing Rosemount 2088 transmitters in the field. See
on page 171
identifies the transient protection terminal block.
Figure 4-10. Transient protection terminal block
A. Lightning bolt location
Note
The transient protection terminal block does not provide transient protection unless the
transmitter case is properly grounded. Use the guidelines to ground the transmitter case. Refer
to
Figure 4-9 on page
Electrical installation
for part numbers. The lightning bolt symbol shown in
48.
Section 4: Electrical Installation
"Spare parts"
Figure 4-10 on page 49