Quick Start - Emerson Rosemount FCL Reference Manual

Free chlorine system with rosemount 1056 transmitter
Hide thumbs Also See for Rosemount FCL:
Table of Contents

Advertisement

Reference Manual
00809-0100-3412
6. Once the cable has been connected to the board, slide the board fully into the
7. Tighten the gland nut to secure the cable and ensure a sealed enclosure.
3.5

Quick Start

Procedure
1. Once connections are secured and verified, apply power to the transmitter.
Rosemount FCL 1056
Connect green wire to metal conduit ground plate in bottom of enclosure.
Figure 3-4: Wiring Diagram for 3900VP-10 pH Sensor (Blue Cable)
A. White
B. White/red
C. Red
D. Blue
E. Clear (not used)
F. Clear
G. Orange
H. White/gray
I. Gray
J. Resistance temperature device return
K. Resistance temperature device sense
L. Resistance temperature device in
M. Ground solution
N. pH shield
O. In pH/ORP
P. Reference shield
Q. In reference
Green (connect to green grounding screw at bottom of enclosure).
enclosure while taking up the excess cable through the cable gland.
When the transmitter is powered up for the first time, Quick Start screens appear.
Using Quick Start is easy.
a. A backlit field shows the position of the cursor.
Wire
May 2019
23