The analyzer probe takes approximately 45 minutes to warm up to the 1,357 °F
(736 °C) heater set point. The 4-20 mA signal remains at a default value of 3.5
mA, and the O
warm-up, the probe begins reading oxygen, and the 4-20 mA output is based on
the default range of 0 to 10 percent O
If there is an error condition at start-up, the Rosemount 6888Xi displays an alarm
message.
5.4 

Run Rosemount 6888Xi Quick Start Wizard

When you first power up the Rosemount 6888Xi, a short wizard program guides
you through the basic setup. Once configured, the Rosemount 6888Xi retains the
setup, and the wizard will not repeat.
Procedure
1. Apply power to the Rosemount 6888Xi.
2. Press Enter to continue.
3. On the Sensor Type screen, use the Up and Down keys to select O
4. Press Enter to continue.
5. On the Device Type screen use the Up and Down keys to select HART
6. At the Auto Cal Device screen, use the Up and Down keys to select the
7. Press Enter to continue.
8. When prompted by Setup Correct? , use the Up and Down keys to select
Manual
reading remains at 0 percent through this warm-up period. After
2
Once the device powers on, the Quick Start Wizard screen appears. With
a dual channel Rosemount 6888Xi, the wizard runs for both channels in
succession.
Do not select CO, as this option is reserved for future use.
(F
Fieldbus), whichever applies.
OUNDATION
calibration method to be used. The methods are defined as follows:
None: Manual calibration with the standard probe housing
configuration.
SPS: Automatic calibration with the standard probe housing
configuration using the Rosemount SPS 4001B.
IMPS: Automatic calibration with the standard probe housing
configuration using the Rosemount IMPS.
Integral: Automatic calibration with the integral autocal probe housing
configuration.
Note
If you select SPS, IMPS, or Integral, you must still configure automatic
calibration as On. Verify other parameters, such as test gas values and
gas times, as well. Refer to the
calibration setup details.
Yes.
.
2
Rosemount 6888Xi Reference Manual
Power up
January 2023
.
2
or FF
®
for
53