Apply Power To The Transmitter And Complete Quick Start - Emerson Rosemount TCL Reference Manual

Total chlorine system with rosemount 56 transmitter
Hide thumbs Also See for Rosemount TCL:
Table of Contents

Advertisement

Wiring
January 2023
6. Once the cable has been connected to the board, slide the board fully into the
7. If you are using a cable gland, tighten the gland nut to secure the cable and ensure
4.7 
Apply power to the transmitter and complete
Quick Start
For Rosemount Total Chlorine System with Rosemount 56 transmitter:
Procedure
1. Once connections are secured and verified, apply power to the transmitter.
2. The next screen lists navigation rules. Press ENTER/MENU for the next screen.
3. Configure sensor 1.
4. Move the cursor to NEXT and press ENTER/MENU.
5. Press ENTER/MENU to show the main display.
22
Figure 4-4: Wiring Power with Standard Cable to Rosemount 1056 Transmitter
enclosure while taking up the excess cable through the conduit opening.
a sealed enclosure.
When the transmitter is powered up for the first time, Quick Start screens appear.
The first Quick Start screen has two control boxes, one for language and the other
for temperature units.
a. The cursor, shown by dark blue backlighting, is on the language control
box. To change the language, press the ENTER/MENU key. A list of available
languages, shown two at a time, appears. Using the Up and Down keys,
scroll (see ) to display the choices. Press ENTER/MENU to select the desired
language. Press Down to move the cursor to the temperature control box. To
change units, press ENTER/MENU and scroll to either °F or °C. Press ENTER/
MENU to store the selection.
b. To move to the next screen, use the navigation keys to move the cursor to
NEXT and press ENTER/MENU.
Sensor 1 is the total chlorine sensor. The screen has two control boxes.
a) For measurement, choose Total chlorine
b) Choose the desired units, mg/L or ppm.
The display changes to show some basic keypad operation guidelines.
The outputs, alarms, display configuration, and data logging are all assigned to
default values. The default value for data logging is disabled.
Reference Manual
00809-0200-3415
Emerson.com/Rosemount