Waters 2998 Overview And Maintenance Manual page 36

Photodiode array detector
Hide thumbs Also See for 2998:
Table of Contents

Advertisement

Table 2–5:
Alliance separations module (connector B)
Pin 4 stop flow (red)
Pin 5 stop flow – (black)
Figure 2–9:
Alliance connector B
Inject Start
Inject Start
Ground
Stop Flow
Stop Flow
Hold Inject 1
Hold Inject 1
Hold Inject 2
Hold Inject 2
Ground
Chart Out
Chart Out
2.6.3.3
Connecting to an Empower or MassLynx data system via an eSAT/IN
module
To send an integrator analog output signal (–0.1 to 2.1 V) from the 2998 PDA detector to an
Empower or MassLynx system (via a two-channel eSAT/IN module), connect cables shown in the
following table and figure.
Table 2–6:
eSAT/IN module connector
CHANNEL 1
2998 PDA detector connections to an Alliance separations module
Stop-flow connections between the Alliance separations module and the
2998 PDA detector
1
2
3
Red
4
5
6
Black
7
8
9
10
11
12
2998 PDA detector connections to the eSAT/IN module
December 5, 2017, 715004753 Rev. B
2998 PDA detector (connector B)
Pin 6 switch 1 (red)
Pin 7 switch 1 – (black)
2998 PDA detector connector B
2998 PDA detector (connector B)
Pin 1 signal out  (white)
Pin 2 signal out – (black)
Page 36
1
Analog 1
2
Analog 1
3
Ground
4
Analog 2
5
Analog 2
6
Switch 1
7
Switch 1
8
Ground
9
Switch 2
10
Switch 2