Waters 2998 Overview And Maintenance Manual page 35

Photodiode array detector
Hide thumbs Also See for 2998:
Table of Contents

Advertisement

To connect the detector to other devices, use the two analog-out/event-in (I/O) connectors and
their mating connectors on the rear panel.
2.6.3.1
Generating an inject-start function
To generate the detector's inject-start function at the start of an injection, connect cables as shown
in the following table and figure.
The firmware defaults to Autozero on injection.
Tip:
Table 2–4:
Alliance separations module (connector B)
Pin 1 inject start (red)
Pin 2 inject start (black)
Figure 2–8:
Alliance connector B
Inject Start
Inject Start
Ground
Stop Flow
Stop Flow
Hold Inject 1
Hold Inject 1
Hold Inject 2
Hold Inject 2
Ground
Chart Out
Chart Out
2.6.3.2
Generating a stop-flow signal
The detector is fitted with a programmable switch output controlled by threshold or timed events.
To enable the generation of a stop flow, connect cables shown in the following table and figure.
To automatically stop the solvent flow within the system in the event of an error
Requirement:
condition or hardware failure, the stop-flow signal must be sent to the system's pump.
2998 PDA detector connections to an Alliance separations module
Inject-start connections between the Alliance separations module and the
2998 PDA detector
Red
1
2
Black
3
4
5
6
7
8
9
10
11
12
December 5, 2017, 715004753 Rev. B
2998 PDA detector (connector A)
Pin 1 inject start  (red)
Pin 2 inject start (black)
2998 PDA detector connector A
Page 35
1
Inject Start
2
Inject Start
3
Ground
4
Lamp On/Off
5
Lamp On/Off
6
Chart Mark
7
Chart Mark
8
Ground
9
Auto Zero
10
Auto Zero