Ulvac G-TRAN Series Instruction Manual page 59

4ch digital display unit
Hide thumbs Also See for G-TRAN Series:
Table of Contents

Advertisement

5.3.3.5. Loading the measurement value/status
Request is made to load the pressure value, status and setpoint information of equipment connected
to each channel.
Request
:
CH
(CH: 1~5)
* The setpoint indicates the allocation of the setpoint output to the designated measuring unit (ch).
* About designation of CH:5
Processed as an effective command only when the combination mode is designated.
The displayed value is transmitted in response as the pressure value.
(In the combination mode, CH:1, 2 is transmitted in response only within the actuating range of
IG2/IG4/CCG or PG.)
Status (SH, SL) reflects the value of the CH1: IG2, IG4 status.
Setpoint status (STSH, STSL) reflects the set status of Ch: 1, 2.
Response
:
CH
'D'
X.XE XX
SH
SL
STSH
STSL
Refer to the next item for more information about status.
[When loading] status list
SH (higher 4 bits)
Bit
Signal identification
No.
B7
FIL-2/1
B6
FIL-ON/OFF
B5
Em.Valid
B4
DEGAS-ON/OFF
B7
FIL-2/1
B6
FIL-ON/OFF
B5
Em.Valid
B4
DEGAS-ON/OFF
With IG
FIL-2/1
FIL-ON/OFF
Em.Valid
DEGAS-ON/OFF
'D'
CR
X
.
X
E
:
Pressure value
:
Higher 4 bits of status
:
Lower 4 bits of status
:
Higher 4 bits of setpoint setting/no setting to ch
:
Lower 4 bits of setpoint setting/no setting to ch
0〔30H〕 4〔34H〕 6〔36H〕 5〔35H〕 7〔37H〕
0
0
0
0
8〔38H〕 C〔43H〕 E〔35H〕 D〔44H〕 F〔46H〕
1
0
0
0
1: FIL-1
1: ON
1: OK
1: ON
±
X
X
SH
Value [ASCII code]
0
0
1
1
0
1
0
0
1
1
1
1
0
1
0
0
0: FIL-2
With CCG, always 0
0: OFF
0: NG
With CCG, always 0
0: OFF
With CCG, always 0
51
SL
STSH
STSL
IG
0
0
1
1
0
1
1
1
1
1
1
1
0
1
1
1
CR