Set-Up And Connection; Location Of Use - Christ Beta 1-8 LSCbasic Operating Manual

Freeze-dryer
Table of Contents

Advertisement

Freeze-dryer Beta 1-8 LSCbasic
Freeze-dryer Beta 2-8 LSCbasic

5 Set-up and connection

5
Set-up and connection
5.1

Location of use

NOTE
36 / 107
Use the freeze-dryer solely in closed and dry spaces.
Refrigeration problems of the freeze-dryer are often caused by insufficient
conditions at the location of use. This is why compliance with the following
conditions is absolutely mandatory!
The table must be stable and have a solid, even tabletop
Ensure sufficient ventilation. Do not place any paper, cloth or similar
material behind or under the unit, since otherwise the air circulation will
be impaired.
Keep a safety distance of at least 30 cm around the freeze-dryer so that
the vents in the unit remain fully effective.
The ambient temperature must be in the range of +5°C to +25°C. A
potential night-time setback of the air conditioning system must be
taken into consideration.
Prevent the room temperature from rising, for example due to closed
doors at night.
Do not subject the freeze-dryer to thermal stress, e.g. by positioning it
near heat generators.
Prevent thermal overload, e.g. caused by other equipment in the direct
vicinity of the freeze-dryer.
Do not set up the vacuum pump in the area of the heat exchanger
ventilation grid (see chapter 2.1.1 - "Functional and operating
elements").
In the case of water-cooled systems, ensure that the water circuit
provides a sufficient amount of cooling water.
Avoid direct sunlight (UV radiation).
Version 03/2018, Rev. 1.5 of 09/03/2020 • sb-dc
Translation of the original operating manual