18.6.5 Channel count
Set the pump channel count to 4.
0xA0004[CR]
*
18.6.6 Roller count
Set the roller count for channel 4 to 6 rollers.
4xB0006[CR]
*
18.6.7 Total Revolutions
Get the total revolutions for channel 3.
3xC[CR]
0000001511[CR][LF]
The total revolution count for channel 3 is 1511.
18.6.8 Volume total
Get the total volume count for channel 3.
3xG[CR]
0000001511[CR][LF]
The total volume count for channel 3 is 1511 mL.
18.6.9 Time total
Get the total pumping time count for channel 3.
3xJ[CR]
0000001511[CR][LF]
The total pumping time count for channel 3 is 1511 seconds.
REGLO ICC - ENGLISH , REV. D, 14-036-E
Get the pump channel count.
REQUEST
RESPONSE
The pump channel count is 3.
Get the roller count for channel 1.
REQUEST
RESPONSE
The roller count for channel 1 is 12.
REQUEST
RESPONSE
REQUEST
RESPONSE
REQUEST
RESPONSE
0xA[CR]
REQUEST
3[CR][LF]
RESPONSE
1xB[CR]
REQUEST
12[CR][LF]
RESPONSE
32 of 36