18.5 Calibration
18.5.1 Volume (target)
Set the target calibration volume for
channel 4 to 100 mL.
4xU1000+2[CR]
1000E+2[CR][LF]
18.5.2 Volume (measured)
Set the measured calibration volume for
channel 4 to 101 mL.
4xV1010+2[CR]
1010E+2[CR][LF]
18.5.3 Time
Set the calibration time for channel 1 to 30 seconds.
1xW00000030[CR]
*
18.5.4 Get time since last calibration
Get the time since channel 3 was last calibrated.
3xX[CR]
03596400[CR][LF]
Channel 3 was last calibrated 3,596,400 seconds (999 hours) ago.
18.5.5 Start calibration
Start calibration on channel 2.
2xY[CR]
*
REGLO ICC - ENGLISH , REV. D, 14-036-E
Get the target calibration volume for channel 3.
REQUEST
RESPONSE
The target calibration volume for channel 3 is 1 mL.
Get the measured calibration volume for channel 3.
REQUEST
RESPONSE
The measured calibration volume for channel 3 is 102 mL.
Get the calibration time setting for channel 3.
REQUEST
RESPONSE
The calibration time for channel 3 is 30 seconds.
REQUEST
RESPONSE
REQUEST
RESPONSE
3xU[CR]
REQUEST
1000E+0[CR][LF]
RESPONSE
3xV[CR]
REQUEST
1020E+2[CR][LF]
RESPONSE
3xW[CR]
REQUEST
30[CR][LF]
RESPONSE