Ismatec MasterFlex Reglo ICC Operating Manual page 28

Table of Contents

Advertisement

18.2.4 Set direction
Set direction to clockwise:
1J[CR]
*
18.3 Modes and Settings
18.3.1 Operational pumping mode
Get the current mode of channel 1.
1xM[CR]
L[CR][LF]
18.3.2 Pumping rate (drive speed)
Set drive speed of channel 2 to 123 RPM:
2S012300[CR]
*
18.3.3 Pumping rate (volume/time)
Set pumping rate of channel 1 to 1.3 μL/min:
1f1300-3[CR]
1300E-3[CR][LF]
18.3.4 Volume
Set the volume of channel 4 to 300 mL:
4v3000+2[CR]
3000E+2[CR][LF]
REGLO ICC - ENGLISH , REV. D, 14-036-E
REQUEST
RESPONSE
The current mode of channel 1 is RPM.
Set the mode of channel 4 to Time.
REQUEST
RESPONSE
Get the current drive speed of channel 4:
REQUEST
RESPONSE
The current drive speed of channel 4 is 12.3 RPM.
Get current pumping rate of channel 4:
REQUEST
RESPONSE
The current pumping rate of channel 4 is 32.5 mL/min.
Get the volume of channel 2:
REQUEST
RESPONSE
The volume of channel 3 is 1.3 μL.
4N[CR]
REQUEST
*
RESPONSE
4S[CR]
REQUEST
12.30[CR][LF]
RESPONSE
4f[CR]
REQUEST
3250E+1[CR][LF]
RESPONSE
2v[CR]
REQUEST
1300E-3[CR][LF]
RESPONSE