In MAX/MIN mode, the operating range is limited to 1kHz.
If the level of the signal is too low (<10% of the range, or V<10V or
I<10A) or if the frequency is less than 5Hz, the device cannot
determine the frequency and displays "----".
Specific Specifications in MAX/MIN mode (from 10Hz to 1kHz):
Accuracy: add 1% R to the values of the table above.
Capture of the extreme: approximately 100ms.

4.2.20 Specifications in THDr

Measurement Range
Specified Measurement
Range
Accuracy
Resolution

4.2.21 Specifications in THDf

Measurement Range
Specified Measurement
Range
Accuracy
Resolution
Note : The display is "----" if the input signal is too low (V<8V or I<9A) or if
the frequency is less than 5Hz.
Specific characteristics in MAX/MIN mode in THD (from 10Hz to 1kHz):
Accuracy: add 1% R to the values in the tables above.
Capture of the extreme: approximately 100ms.
GlobalTestSupply
www.
0.0 to 100%
0 to 100%
of the measurement range
± (5% R ±2cts) in voltage
± (5% R ±5cts) in current
0.1%
0.0 to 1000%
0 to 100%
of the measurement range
± (5% R ±2cts) in voltage
± (5% R ±5cts) in current
0.1%
45