3.11 POWER MEASUREMENTS W, VA, VAR AND PF
This measurement is possible in single-phase or in balanced 3-phase.
NOTE: If performing DC or AC+DC power measurements, correct the DC zero in
current first (see § 3.9.2)
For the power factor (PF) and the powers VA and var, the measurement is
possible only in AC or AC+DC.
3.11.1 Single-Phase Power Measurement
1.
Set the switch to
button until the desired choice is reached.
2.
The device automatically displays AC+DC. To select AC, DC, or
AC+DC, press the
reached.
3.
Connect the black lead to the COM terminal and the red lead to the "+"
terminal.
4.
Connect the test probes or the alligator clips of the black lead on the
neutral (N), then those of the red lead on the L phase.
5.
Clamp around only the corresponding conductor, respecting the
direction.
The measurement is displayed on screen.