Chapter 8 Other Features; Selecting The Resolution Mode - YOKOGAWA 2560A User Manual

Precision dc calibrator
Hide thumbs Also See for 2560A:
Table of Contents

Advertisement

Chapter 8
Other Features
8.1

Selecting the Resolution Mode

Procedure
1.
Turn the RANGE dial to select SETUP.
The SETUP menu appears.
1. Select SETUP.
2.
Turn a main setting dial to select RESOL.
The deviation display shows the current resolution mode.
3.
Turn a deviation dial to select NORM or HIGH.
When you change the setting, all PRESET indicators turn on.
If you do not change the setting, the PRESET indicators will not turn on.
4.
Flip the PRESET switch down.
The setting is confirmed. All the PRESET indicators turn off.
The 2560A returns from the SETUP menu to source mode, and the range is set to 1 V.
IM 2560A-01EN
The SETUP indicator turns on.
RANGE dial
Main setting dials (use any)
2. Select RESOL.
Deviation dials (either)
4. Confirm the settings.