Selecting The Temperature Unit - YOKOGAWA 2560A User Manual

Precision dc calibrator
Hide thumbs Also See for 2560A:
Table of Contents

Advertisement

6.7

Selecting the Temperature Unit

Procedure
1.
Turn the RANGE dial to select SETUP.
The SETUP menu appears.
1. Select SETUP.
The SETUP indicator turns on.
2.
Turn a main setting dial to select TEMP.
"NEXT" appears on the deviation display. All the PRESET indicators turn on.
3.
Flip the PRESET switch down.
The TEMP menu appears. All the PRESET indicators turn off.
4.
Turn a main setting dial to select UNIT.
The deviation display shows the current temperature unit.
5.
Turn a deviation dial to select CEL or FAR.
When you change the setting, all PRESET indicators turn on.
If you decide not to change the setting, step 6 is not necessary.
6.
Flip the PRESET switch down.
The setting is confirmed. All the PRESET indicators turn off.
-
IM 2560A-01EN
RANGE dial
Main setting dials (use any)
2. Select TEMP.
Main setting dials (use any)
4. Select UNIT.
PRESET indicator
3. Display the TEMP menu.
PRESET indicator
Deviation dials (either)
5. Select CEL or FAR.
6. Confirm the settings.
1
2
3
4
5
6
7
8
9
10
11
PRESET switch
12
13
14
15
16
App
PRESET switch
Index
6-19