15.10 Cleaning the Filter

If the filter attached to the air inlet is dirty, the air intake may be obstructed causing the
instrument to heat up and break down.
French
Si le filtre raccordé aux orifices d'entrée est poussiéreux, l'admission d'air risque de s'obstruer,
ce qui risque d'entraîner une surchauffe de l'équipement et sa panne.
There is a filter attached to the air inlet on the rear panel of the 2560A.
Check the filter periodically to see that it is not clogged.
Removing the Filter
1.
Turn off the 2560A.
2.
Pull the lock buttons at the left and right edges of the filter cover toward you.
Pull them until you hear a click. This will release the lock.
3.
Pull the filter cover toward you.
The filter cover will come off.
4.
Remove the filter from the filter cover.
15-18
ATTENTION
Lock buttons
Filter
Filter cover
CAUTION
IM 2560A-01EN