Advertisement

5.4

Turning the Output On and Off

• Before turning on the output of the 2560A, make sure that wiring is connected properly.
• Do not touch the terminals or cables when the 2560A output is on. Doing so may cause
electric shock.
French
• Avant de mettre sous tension la sortie du dispositif 2560A, assurez-vous que le câblage est
correctement raccordé.
• Ne touchez jamais les bornes ou les câbles lorsque la sortie du dispositif 2560A est sous
tension. Ceci risquerait d'entraîner un choc électrique.
Procedure
Flip the OUTPUT switch up.
The output turns on. The OFF indicator turns off, and the ON indicator turns on.
Flipping the OUTPUT switch down while the output is on will turn off the output. The ON indicator will
turn off, and the OFF indicator will turn on.
Note
When outputting current, do not open the output terminals.
When outputting voltage, do not short the output terminals.
In either case, the overload protection function will be activated, an error code will be displayed, and the
output will be turned off. The time it takes for the error to appear depends on the range and output value.
If the output terminal temperature changes as a result of making contact with the lead wires, crimping
terminals, banana plugs, or other conductors or due to ambient air movement, errors exceeding the
accuracy specifications may occur when a small output range is used. In such situations, wait until the
output stabilizes. Also while the 2560A is generating output, keep the output terminal temperature from
changing due to the effects air conditioning, other heat sources, and the like.
IM 2560A-01EN
WARNING
AVERTISSEMENT
Output display (output display example for +100.000 V)
OUTPUT switch
Unit
5-7
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index