Setting The Output Scaling - YOKOGAWA 2560A User Manual

Precision dc calibrator
Hide thumbs Also See for 2560A:
Table of Contents

Advertisement

5.3

Setting the Output Scaling

Procedure
Specifying the Main Setting (MAX)
1.
Flip the SCALE switch to the MAX side.
2.
Turn the main setting dials to specify the main setting (MAX).
Specifying the Main Setting (MIN)
3.
Flip the SCALE switch to the MIN side.
4.
Turn the main setting dials to specify the main setting (MIN).
Setting the minimum value to a value other than ±0 (zero) causes the Non Zero indicator to turn on.
Example of setting +10.0000 V
Explanation
Scaling is a feature used to set the maximum value (main setting (MAX)) and minimum value (main
setting (MIN)) of the output range to your liking within the source range of each range. Output divider,
deviation, and sweep features operate within this scale range. Scaling cannot be used in high
resolution mode.
Specifying the Main Setting (MAX) and Main Setting (MIN)
You can set these values within the range that meets the following conditions.
• The main setting (MAX) must be greater than or equal to the main setting (MIN).
• If both the main setting (MAX) and main setting (MIN) are negative, the absolute value of the main
setting (MAX) must be greater than or equal to that of the main setting (MIN).
• You cannot set the main setting (MAX) to a negative value and the main setting (MIN) to a positive
value.
• For details on setting the main setting (MAX) and main setting (MIN) and the resulting conditions,
see the following table.
Original Polarities of the
Main Setting (MAX) and
Main Setting (MIN)
Both positive
IM 2560A-01EN
Main setting dials
2. Set main setting (MAX).
4. Set main setting (MIN).
Setup Operation
Set MAX less than or equal to MIN.
Set MIN greater than or equal to MAX
Set MAX to negative
with the POLARITY switch
Set MIN to negative
with the POLARITY switch
SCALE switch
1. Flip to the MAX side.
3. Flip to the MIN side.
Resulting Condition
MIN = MAX (the specified MAX value)
MAX=MIN (the specified MIN value)
MIN is also set to negative.
MAX stays positive.*
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
5-5

Advertisement

Table of Contents
loading

Table of Contents