Other Features - YOKOGAWA 2560A User Manual

Precision dc calibrator
Hide thumbs Also See for 2560A:
Table of Contents

Advertisement

2.8

Other Features

High Resolution Mode
Normally, set the main setting of the source voltage or current at 5.5 digit resolution (4.5 digit resolution
for 30 A range). In high resolution mode, you can set it at 6.5 digit resolution (5.5 digit resolution for 30
A range).
In normal resolution mode, use the five main setting dials to set 5.5 digits. In high resolution mode, use
the five main setting dials and deviation dial 1 to set 6.5 digits.
In high resolution mode, the output divider, deviation output, sweep, and scaling features cannot be
used. Further, high resolution mode cannot be used for temperature or resistance ranges.
Turning the Beep Sound On and Off
Beeps are used to indicate the operation status of the device. You can turn it on and off from the
SETUP menu.
You can turn the beep sound on or off for the following notifications. You cannot change the volume.
• When a voltage greater than or equal to +150 V or less than or equal to –150 V is to be output,
regardless of the output on/off state
• When a sweep operation is held
You cannot turn off the beep sound for the following notifications.
• When the cooling fan stops (error code: E.901, E.907)
• When an abnormal temperature is detected (error code: E.902, E.903)
• When an internal power supply error is detected (error code: E.904, E.905, E.908)
• When an error is detected during a self-test executed at power-on or by some other operation (error
codes: E.911 to E950). However, you can turn off the beep sound for these errors by performing the
procedure below.
• Turn the power on
Control a front-panel dial or switch
• Execute a self-test using other controls
Flip the PRESET switch down
Error Log Display
The error log keeps a record of error codes that occur in communication and self-tests while the 2560A
is on. You use the SETUP menu to display the error log. The error log is cleared when the power is
turned off.
Initialization
This feature initializes the 2560A settings. You use the SETUP menu to do so.
Communication settings and a portion of other settings are not initialized. For details on the items that
are initialized, see appendix 7.
Displaying the Product Information
The firmware version, serial number, and the like can be displayed. You use the SETUP menu to
display them.
• Firmware version
• Logic program version
• Boot program version
• Serial number (instrument number)
2-12
IM 2560A-01EN