Specifications And Characteristics - Keysight N2818 User Manual

Differential probes
Table of Contents

Advertisement

Specifications and Characteristics

Description
Bandwidth (–3 dB)
Attenuation Ratio
Probe Risetime (10% – 90%)
Gain Accuracy (% of reading)
Absolute Maximum Rated Input Voltage (each
side to ground)
Maximum Differential Input Voltage
(DC + AC Peak)
Maximum Common Mode Input Voltage
Input Resistance ǁ Capacitance
Each Side to Ground:
Between Inputs:
Output Voltage Swing
Offset (typical)
Offset Adjustment Range
AC CMRR
Noise Referenced to Input, Probe Only
Power Requirements
*
All entries are typical unless otherwise noted.
Warranted Specification.
Keysight N2818/9A Differential Probes User Guide
The following tables list the characteristics for the N2818A and N2819A
differential probes. Before any testing and to allow the probe to warm up,
connect the probe to a powered-on oscilloscope for at least 20 minutes.
Ensure that the environmental conditions do not exceed the probe's
specified limits.
Table 6
N2818A without the
N7013/14A Extreme
Temperature Probing Kit
200 MHz
10:1
1.75 ns
±
2%
±
60V
±
20V
±
60V
500 kΩ ǁ 7 pF
1 MΩ ǁ 3.5 pF
±
2V (driving 50Ω scope input)
±
2 mV
–95 mV to +95 mV
> –80 dB at 60 Hz
> –50 dB at 10 MHz
6 mV
rms
6 mV
rms
AutoProbe Interface
Electrical Specifications and Characteristics
N2818A with the N7013/14A
Extreme Temperature Probing
*
Kit
70 MHz
10:1
1.75 ns
±
2%
±
60V
±
20V
±
60V
500 kΩ ǁ 7 pF
1 MΩ ǁ 3.5 pF
±
2V (driving 50Ω scope input)
±
2 mV
–95 mV to +95 mV
> –80 dB at 60 Hz
> –50 dB at 10 MHz
6 mV
rms
6 mV
rms
AutoProbe Interface
Using the Probe
N2819A
800 MHz
10:1
437 ps
±
2%
±
40V
±
15V
±
30V
100 kΩ ǁ 2 pF
200 kΩ ǁ 1 pF
±
1.5V (driving 50Ω scope
input)
±
5 mV
–20 mV to +20 mV
> –60 dB at 60 Hz
> –15 dB at 500 MHz
4.7 mV
rms
4.7 mV
rms
AutoProbe Interface