Characteristics And Specifications - Keysight N2894-60002 User Manual

Passive probe
Table of Contents

Advertisement

Characteristics and Specifications

This section lists the characteristics and specifications for the probes. The probe
and oscilloscope should be warmed up for at least 20 minutes before any testing
and the environmental conditions should not exceed the probe's specified limits.
Table 3 Electrical and Mechanical Characteristics
Description
Attenuation ratio
Bandwidth (–3 dB)
Probe Risetime (10%-90%)
Maximum Rated Input Voltage
Input Resistance (scope + probe)
Input Capacitance (system)
Compensation Range
Input Coupling of the Measuring Instru-
ment
Weight (probe only)
Cable Length
Probe Barrel Diameter
a.700 MHz BW only available on DSOX/MSOX 4000A-series oscilloscopes with 1 GHz or 1.5 GHz
bandwidth
b.300 Vrms, 400 V (DC + peak AC) mains isolated; 0 V transient overvoltage with InfiniiVision 6000X
c.Maximum pulse rating = 1,650 V (step 0 V to 1,650 V), mains isolated, no overshoot permitted,
connected to oscilloscope with an input impedance of 1 M
InfiniiVision 6000X.
d.Measurement Category II
Table 4 Safety Specifications
Specification
CEI/IEC 61010-031
Characteristics and Specifications
Characteristic
10:1
a
700 MHz
500 ps
400 V mains isolated
Ω
10 M
9.5 pF
10 - 25 pF
Ω
1 M
48 g
1.3 m
2.5 mm
Ω
± 1%, not applicable with
9
,
b
c
d
, 300 V CAT II

Advertisement