Minimum Bg For Calcs - Insulet Omnipod Dash User Manual

Insulin management system
Table of Contents

Advertisement

2 Initial PDM Setup
Review the Target BG and Correct
Above values
1. Review the segments for the full 24-
hour profile.
2. To change any of the entries:
a. Tap the row containing the entry to
be changed and enter the corrected
value.
b. Review and correct as needed any
remaining segments.
3. When the segments and values are
correct, tap SAVE.

Minimum BG for Calcs

If your blood glucose level is below the
Minimum BG for Calcs that you specify,
the Bolus Calculator is disabled (see "The
Bolus Calculator" on page 143).
1. To enter your Minimum BG for Calcs,
tap the Min BG for Calc field and enter
the number on the number pad.
Tip: Alternatively, touch the blue circle
and slide it to the desired number.
2. Tap NEXT.