12 Understanding PDM and Pod Function
You may have different routines on different days of the week; for example
your weekend routine may differ from your weekday routine. To handle these
predictable changes in your routine, you can create up to 12 different Basal
Programs (see "Basal Programs" on page 75). This figure shows two possible Basal
Programs, one for weekdays and the other for weekends.
1.0
0.8
0.6
0.4
0.2
0
Midnight

Temporary Basal Rates

The ability to set temporary basal rates, also called "temp basals",
is an important feature of the Omnipod DASH™ System. A temp basal lets
you override the currently running Basal Program by setting a different basal
rate for a predetermined period of time.
For example, if you are going cross-country skiing for several hours, you
could set a temp basal to lower your basal rate during and after your exercise
(see "Using Temporary Basal Rates" on page 69).
Temp basals can last from 30 minutes to 12 hours. At the end of the specified time,
the Pod automatically reverts to the active Basal Program.
Temp basal settings: Units per hour (U/hr) or per cent (%)
Temp basals can be set using percentage (%) or units per hour (U/hr).
Setting temp basals to units per hour (U/hr) means that the Pod delivers insulin
at a flat rate for the duration of the temp basal. In other words, the details of the
currently scheduled Basal Program are ignored during these temp basals.
Setting temp basals to per cent (%) means that insulin delivery follows the pattern
defined by the currently scheduled Basal Program, but increases or decreases the
insulin delivery by the specified percentage. For example, a 50% increase raises
the Basal Program' s insulin delivery by 50%, while a 50% decrease lowers the Basal
Program' s insulin delivery by 50%.
138
Weekday Basal Program
Weekend Basal Program
06:00
Noon
18:00
Midnight