Note: You cannot set a Maximum Basal Rate that is lower than the highest basal
rate of an existing Basal Program, temp basal preset or currently running temp
basal.

Temp basal

To turn on or off the ability to set temp basals:
1. Navigate to: Settings icon (
2. To enable or disable the ability to set temporary basal rates (temp basals), tap
the toggle on or off.
3. To change between using per cent (%) or flat rate (U/hr) temp basals:
a. Tap Temp Basal.
b. Select the desired method for setting a temp basal:
• Tap Per cent (%) to modify the active Basal Program by a set
percentage increase or decrease.
• Tap Flat Rate (U/hr) to replace the active Basal Program with a fixed
basal rate for the specified duration.
c. Tap SAVE.

Bolus Delivery Settings

These settings allow you to change your Maximum Bolus, Extended Bolus and
Bolus Calculator settings.

Maximum Bolus

The Maximum Bolus defines the upper limit for a bolus. The Bolus Calculator
informs you if it calculates a bolus that is over this amount. The largest allowed
value of the Maximum Bolus is 30 units.
Caution: Check with your healthcare provider before adjusting this setting.
To change your Maximum Bolus:
1. Navigate to: Settings icon (
2. Tap the Max Bolus field and enter the new Maximum Bolus value.
3. Tap SAVE.
Adjusting Settings 9
) > Basal & Temp Basal.
) > Bolus > Maximum Bolus.
Effective Date: 27SEP2019, ECO-000240